Home | About us| Products| Order Now| Contact us| Site map|


Air Sterilizer(AS-1)
Air Sterilizer(AS-2)
Blood Bank Refrigerators
Colony Counter
Cooled Incubater
Differential Blood Cell Counter
Dry-Baths
Freezer -30 ํ C
Haema Mixer
Lamp
Ozone Generator
Pharmaecutical Refrigerators
Power Supply For Electrophoresis
Precision Tachometer
Precision Timer
Slide Warmer
Temperature Controller
Universal programable timer
UV/Black/White Trans-illuminator(UVBW-1)
UV/Black/White Trans-illuminator(UVBW-2)
Variable Speed Rotator
Water Circulationg Bath
Water Container

SPECIFICATIONS  
Output range / resolution 0 - 500 / 1 volts
0 -100 / 1 mA
0.1 - 50 / 0.1 watts*
0-99 hr / 1 sec.
Input supply 220 Volt 50 Hz.
Power Supply For Electrophoresis
Model PS-5001 , PS-5001 PT

500 Volts 100 mA 50 Watts*
Constant voltage
Constant current
Constant power or temperature limitation*
Integrator - Timer

CONTROL OF TEMPERATURE*

A temperature probe is connected to the chamber Then a temperature limit can be preset instead of a power limit (0o to 99o). During the run, as soon as the preset temperature is reached, the output voltage will decrease automatically until the temperature of the chamber is lower than preset temperature. The cycle is repeated during the run until the automatic preset shutdown of the timer.

FEATURES
Eight programs with automatic preset cut off. The menu driven LCD back light displayed with 4 lines of 20 characters. When output supply stops, a specific message is displayed. Four operating modes:
- Constant voltage
- Constant current
- Constant power*
- Temperature limitation.*
It is possible to perform electrophoresis with a constant parameter whilst limiting the others to selected values, including temperature.*
Automatic restart in case of power failure recommended for over night electrophoresis. Automatic storage of the preset and elapsed values in case of power failure or voluntary interruption during the cycle.This microprocessor controlled power supply is equip with safety feature a circuit breaker which automatically cut the electrical supply in case of ground leakage detection, short circuit, or overload.
*PS-5001 PT ONLY

© Copyright WorldMedic Corporation Inc. All Rights Reserved.1999-2005